Class Muovi
A BaseBlock that streams high-density surface EMG (HD-sEMG), EEG, or IMU data from one or more Muovi probes connected through an OT Bioelettronica SyncStation.
JSON configuration
Use these block entries inside a pipeline configuration. Input names must match other blocks in that pipeline; device addresses and file paths must match your setup.
Minimal JSON configuration streaming two probes in EMG mode with IMU:
{
"Name": "HD-EMG",
"Type": "Muovi",
"DesiredRate": 100,
"Params": [ "1", "3", "emg" ]
}
Params accepts zero-based probe indices and optional keywords:
"emg", "emg_lowgain" / "lowgain", "eeg", and "imu".
If no indices are specified, probe 0 is used by default.
How Type, Inputs and Params work · Block catalogue and parameter reference
Implements
Inherited Members
Namespace: MOSAIC.Models.Devices
Assembly: MOSAIC.dll
Syntax
public sealed class Muovi : BaseBlock, INotifyPropertyChanged, INotifyPropertyChanging, ISubscriber, IPublisher, IBaseBlockProps, IJsonExportable, IDisposable, IAsyncDisposable
Remarks
The Muovi probe is a wireless, wearable, high-density biosignal acquisition device
manufactured by OT Bioelettronica (Turin, Italy). Each probe features
a 32-electrode grid and a 9-axis IMU. Up to four probes can be synchronised through
a single SyncStation base unit, which exposes a TCP server at 192.168.76.1:54320.
Firmware documentation and protocol specifications are available from the manufacturer at https://otbioelettronica.it/en/download/#55-171-wpfd-muovi.
Supported signal modes:
| Mode | Sample rate / gain |
|---|---|
| Emg | 2 000 Hz, gain ×8 (286.1 nV/bit, ±9.375 mV range) |
| EmgLowGain | 2 000 Hz, gain ×4 (572.2 nV/bit, ±18.75 mV range) |
| Eeg | 500 Hz, EEG-optimised front-end |
Output modes:
-
Matrix push — when DesiredRate is lower than
the nominal sample rate, the reader thread accumulates
NominalRate / DesiredRatesamples and publishes a MathNet.Numerics.LinearAlgebra.Matrix<T> directly; no upstream timer is required. - Vector timer — when DesiredRate equals the nominal rate, the reader decodes every sample and a timer tick publishes the latest MathNet.Numerics.LinearAlgebra.Vector<T>.
Protocol overview (SyncStation v2.8):
The configuration packet consists of a START byte encoding the number of CONTROL bytes
that follow (one per enabled probe), plus a CRC-8 trailer. The SyncStation allocates
frame bandwidth only for probes whose CONTROL bytes are present, so the total raw columns
per sample equals enabledProbes × 38 + 6 (38 channels per probe plus 6 SyncStation
accessory channels). Each TCP burst contains exactly 18 consecutive samples.
Constructors
Muovi(string, double)
Initialises a new instance of the Muovi class with default settings.
Declaration
public Muovi(string name = "Muovi", double desiredRate = 100)
Parameters
| Type | Name | Description |
|---|---|---|
| string | name | Display name for this block instance. |
| double | desiredRate | Desired output rate in Hz. If lower than the probe's nominal rate, multiple samples are batched into a MathNet.Numerics.LinearAlgebra.Matrix<T>. If equal, single-sample MathNet.Numerics.LinearAlgebra.Vector<T> mode is used. |
Fields
SubSamplingRate
Number of samples the SyncStation packs into each TCP burst. This value is fixed in firmware and cannot be changed.
Declaration
private const int SubSamplingRate = 18
Field Value
| Type | Description |
|---|---|
| int |
SyncStationExtraCh
Number of accessory channels appended by the SyncStation after all probe data.
Declaration
private const int SyncStationExtraCh = 6
Field Value
| Type | Description |
|---|---|
| int |
SyncStationIp
Default IP address of the SyncStation Wi-Fi access point.
Declaration
private const string SyncStationIp = "192.168.76.1"
Field Value
| Type | Description |
|---|---|
| string |
SyncStationPort
TCP port exposed by the SyncStation firmware.
Declaration
private const int SyncStationPort = 54320
Field Value
| Type | Description |
|---|---|
| int |
_bytesPerPacket
Declaration
private int _bytesPerPacket
Field Value
| Type | Description |
|---|---|
| int |
_bytesPerSample
Declaration
private int _bytesPerSample
Field Value
| Type | Description |
|---|---|
| int |
_client
Declaration
private TcpClient? _client
Field Value
| Type | Description |
|---|---|
| TcpClient |
_conversionFactor
Declaration
private double _conversionFactor
Field Value
| Type | Description |
|---|---|
| double |
_hasEmg
Indicates whether electrode data is available (always true while streaming).
Declaration
[ObservableProperty]
private bool _hasEmg
Field Value
| Type | Description |
|---|---|
| bool |
_hasImu
Indicates whether IMU data is available (mirrors StreamImu while streaming).
Declaration
[ObservableProperty]
private bool _hasImu
Field Value
| Type | Description |
|---|---|
| bool |
_isStreaming
Declaration
private volatile bool _isStreaming
Field Value
| Type | Description |
|---|---|
| bool |
_lastRamp
Declaration
private short _lastRamp
Field Value
| Type | Description |
|---|---|
| short |
_latestImu
Declaration
private volatile Vector<double>? _latestImu
Field Value
| Type | Description |
|---|---|
| Vector<double> |
_latestSample
Declaration
private volatile Vector<double>? _latestSample
Field Value
| Type | Description |
|---|---|
| Vector<double> |
_matrixMode
Declaration
private bool _matrixMode
Field Value
| Type | Description |
|---|---|
| bool |
_nominalRate
Declaration
private int _nominalRate
Field Value
| Type | Description |
|---|---|
| int |
_rampInitialized
Declaration
private bool _rampInitialized
Field Value
| Type | Description |
|---|---|
| bool |
_readerThread
Declaration
private Thread? _readerThread
Field Value
| Type | Description |
|---|---|
| Thread |
_reading
Declaration
private volatile bool _reading
Field Value
| Type | Description |
|---|---|
| bool |
_sampleReady
Declaration
private volatile bool _sampleReady
Field Value
| Type | Description |
|---|---|
| bool |
_samplesPerBatch
Declaration
private int _samplesPerBatch
Field Value
| Type | Description |
|---|---|
| int |
_stream
Declaration
private NetworkStream? _stream
Field Value
| Type | Description |
|---|---|
| NetworkStream |
_totalRawColumns
Declaration
private int _totalRawColumns
Field Value
| Type | Description |
|---|---|
| int |
Properties
ChannelLabels
All output channel labels in order (EMG/EEG then IMU, repeated per probe).
Declaration
public List<string> ChannelLabels { get; private set; }
Property Value
| Type | Description |
|---|---|
| List<string> |
ChannelsPerProbeOutput
Number of output channels per probe: 32 electrode channels, plus optionally 4 IMU channels.
Declaration
public int ChannelsPerProbeOutput { get; }
Property Value
| Type | Description |
|---|---|
| int |
DumpFilePrefix
Recordings from this block are named <block>_muovi.csv.
Declaration
protected override string DumpFilePrefix { get; }
Property Value
| Type | Description |
|---|---|
| string |
Overrides
EmgChannelLabels
Electrode-only channel labels, used to populate the EMG/EEG scope legend.
Declaration
public List<string> EmgChannelLabels { get; private set; }
Property Value
| Type | Description |
|---|---|
| List<string> |
ForceMatrixMode
Explicit override for the publish strategy, set from the JSON "matrix" /
"vector" param. When null (default), the mode is auto-derived
from DesiredRate vs the probe's nominal rate (legacy behaviour):
batch into a MathNet.Numerics.LinearAlgebra.Matrix<T> when more than one sample falls in an output period,
otherwise publish single MathNet.Numerics.LinearAlgebra.Vector<T> samples. When set, it forces the mode:
- true — always publish a per-packet MathNet.Numerics.LinearAlgebra.Matrix<T> (every acquired sample is kept; correct for high-rate display/recording).
- false — always publish decimated single MathNet.Numerics.LinearAlgebra.Vector<T> samples at DesiredRate.
Must be set before Connect().
Declaration
public bool? ForceMatrixMode { get; set; }
Property Value
| Type | Description |
|---|---|
| bool? |
HasEmg
Indicates whether electrode data is available (always true while streaming).
Declaration
[ExcludeFromCodeCoverage]
public bool HasEmg { get; set; }
Property Value
| Type | Description |
|---|---|
| bool |
HasImu
Indicates whether IMU data is available (mirrors StreamImu while streaming).
Declaration
[ExcludeFromCodeCoverage]
public bool HasImu { get; set; }
Property Value
| Type | Description |
|---|---|
| bool |
ImuChannelLabels
IMU-only channel labels, used to populate the IMU scope legend.
Declaration
public List<string> ImuChannelLabels { get; private set; }
Property Value
| Type | Description |
|---|---|
| List<string> |
IsStreaming
Indicates whether the block is currently connected to the SyncStation and streaming data.
Declaration
public bool IsStreaming { get; }
Property Value
| Type | Description |
|---|---|
| bool |
MaxInputs
Gets the maximum number of input connections allowed for this block.
Declaration
public override int MaxInputs { get; }
Property Value
| Type | Description |
|---|---|
| int |
Overrides
Remarks
Override in derived classes to specify input requirements. Default is 1 (most blocks accept exactly one input). Use MaxValue for blocks that accept unlimited inputs (e.g., Joiner).
MinInputs
Gets the minimum number of input connections required for this block.
Declaration
public override int MinInputs { get; }
Property Value
| Type | Description |
|---|---|
| int |
Overrides
Remarks
Override in derived classes to specify input requirements. Default is 1 (most blocks require exactly one input). Source blocks (devices, generators) should return 0.
OutputChannels
Total number of output channels across all enabled probes.
Equals ProbeIndices.Count × ChannelsPerProbeOutput.
Declaration
public int OutputChannels { get; }
Property Value
| Type | Description |
|---|---|
| int |
ProbeIndices
Zero-based indices of the Muovi probes to enable on the SyncStation. The index corresponds to the physical slot on the SyncStation (0–3 for standard Muovi probes, 4–15 for Muovi+ and other OT Bioelettronica devices). Verify the correct index against the LED indicators on the SyncStation.
Declaration
public List<int> ProbeIndices { get; }
Property Value
| Type | Description |
|---|---|
| List<int> |
ScopeEmg
Scope monitor for real-time visualisation of the electrode (EMG / EEG) channels.
Bind to a ScopeMonitorView in the UI layer.
Declaration
public ScopeMonitor ScopeEmg { get; }
Property Value
| Type | Description |
|---|---|
| ScopeMonitor |
ScopeImu
Scope monitor for real-time visualisation of the IMU quaternion channels. Only populated when StreamImu is true.
Declaration
public ScopeMonitor ScopeImu { get; }
Property Value
| Type | Description |
|---|---|
| ScopeMonitor |
SignalMode
Gets or sets the signal acquisition mode for all enabled probes. Must be set before calling Connect().
Declaration
public MuoviSignalMode SignalMode { get; set; }
Property Value
| Type | Description |
|---|---|
| MuoviSignalMode | The default is Emg. |
StreamImu
Gets or sets whether the four IMU quaternion channels (W, X, Y, Z) per probe are included in the output vector. Must be set before calling Connect().
Declaration
public bool StreamImu { get; set; }
Property Value
| Type | Description |
|---|---|
| bool |
StreamInfo
Observable telemetry counters (frames collected, packets lost, pipeline status) suitable for binding in a device card view.
Declaration
public MuoviStreamInfo StreamInfo { get; }
Property Value
| Type | Description |
|---|---|
| MuoviStreamInfo |
Methods
BuildChannelLabels()
Rebuilds ChannelLabels, EmgChannelLabels, and ImuChannelLabels based on the current ProbeIndices, SignalMode, and StreamImu settings.
Declaration
private void BuildChannelLabels()
CheckRamp(byte[], int)
Validates the per-sample ramp counter embedded in the last raw column of each frame. Sets DataLoss when a gap is detected and resets it when consecutive clean samples arrive.
Declaration
private void CheckRamp(byte[] frame, int offset)
Parameters
| Type | Name | Description |
|---|---|---|
| byte[] | frame | The raw byte buffer. |
| int | offset | Byte offset of the current sample within |
Remarks
The SyncStation ramp counter increments by 256 per firmware packet (not per sample). A gap of 257 at the 16-bit wrap boundary is normal; gaps ≥ 258 indicate packet loss.
ConfigureInput(IServiceProvider, JsonModel)
Factory method that creates and configures a Muovi instance from a JSON pipeline definition.
Declaration
public static Muovi ConfigureInput(IServiceProvider sp, JsonModel m)
Parameters
| Type | Name | Description |
|---|---|---|
| IServiceProvider | sp | The application's dependency injection service provider. |
| JsonModel | m | The JSON model containing |
Returns
| Type | Description |
|---|---|
| Muovi | A fully configured Muovi block ready to be connected. |
Remarks
Recognised Params values:
- Integer values — interpreted as zero-based probe slot indices.
"emg"— standard EMG mode (default)."emg_lowgain"or"lowgain"— EMG with reduced gain."eeg"— EEG acquisition mode at 500 Hz."imu"— include IMU quaternion channels in output.
Connect()
Opens a TCP connection to the SyncStation, sends the configuration packet, and starts the background reader thread.
Declaration
public void Connect()
Remarks
Set SignalMode, StreamImu, and ProbeIndices before calling this method. The SyncStation begins transmitting data immediately after the configuration packet is acknowledged. The method is a no-op if the block is already streaming — call Disconnect() first to reconfigure.
Exceptions
| Type | Condition |
|---|---|
| SocketException | Thrown if the TCP connection to the SyncStation cannot be established (e.g. device not powered on or Wi-Fi not connected). |
ConnectAsync()
Asynchronously connects to the SyncStation on a background thread.
Declaration
public Task ConnectAsync()
Returns
| Type | Description |
|---|---|
| Task | A task that completes when the connection is established and streaming has started. |
DecodeSampleInline(byte[], int)
Decodes a single sample from the raw byte frame into the latest MathNet.Numerics.LinearAlgebra.Vector<T> and optional IMU vector (vector / timer mode).
Declaration
private void DecodeSampleInline(byte[] frame, int offset)
Parameters
| Type | Name | Description |
|---|---|---|
| byte[] | frame | The accumulation buffer containing raw bytes. |
| int | offset | Byte offset of the sample within |
Disconnect()
Stops the reader thread, sends a shutdown packet to the SyncStation, and closes the TCP connection.
Declaration
public void Disconnect()
Remarks
Safe to call even if the block is not currently streaming (no-op in that case).
The shutdown packet (0x00 + CRC) instructs the SyncStation firmware to
stop transmitting.
Dispose()
Releases all resources: pauses and disposes both scope monitors, disconnects from the SyncStation, and flushes the CSV dumper.
Declaration
public override void Dispose()
Overrides
ExtractElectrode(Vector<double>)
Extracts only the electrode channels from a full output vector that may contain interleaved IMU channels, producing a contiguous EMG/EEG-only vector for ScopeEmg.
Declaration
private Vector<double> ExtractElectrode(Vector<double> full)
Parameters
| Type | Name | Description |
|---|---|---|
| Vector<double> | full | The full output vector including both electrode and IMU channels. |
Returns
| Type | Description |
|---|---|
| Vector<double> | A vector containing only electrode channels. If StreamImu is
false, returns |
ExtractElectrodeMatrix(Matrix<double>)
Matrix counterpart of ExtractElectrode(Vector<double>): returns a [rows × (probes·electrodeChannels)] sub-matrix containing only the electrode columns, stripping the interleaved IMU columns. When StreamImu is false the matrix is returned unchanged (no copy).
Declaration
private Matrix<double> ExtractElectrodeMatrix(Matrix<double> full)
Parameters
| Type | Name | Description |
|---|---|---|
| Matrix<double> | full |
Returns
| Type | Description |
|---|---|
| Matrix<double> |
OnHasEmgChanged(bool)
Executes the logic for when HasEmg just changed.
Declaration
private void OnHasEmgChanged(bool value)
Parameters
| Type | Name | Description |
|---|---|---|
| bool | value | The new property value that was set. |
Remarks
This method is invoked right after the value of HasEmg is changed.
OnHasEmgChanged(bool, bool)
Executes the logic for when HasEmg just changed.
Declaration
private void OnHasEmgChanged(bool oldValue, bool newValue)
Parameters
| Type | Name | Description |
|---|---|---|
| bool | oldValue | The previous property value that was replaced. |
| bool | newValue | The new property value that was set. |
Remarks
This method is invoked right after the value of HasEmg is changed.
OnHasEmgChanging(bool)
Executes the logic for when HasEmg is changing.
Declaration
private void OnHasEmgChanging(bool value)
Parameters
| Type | Name | Description |
|---|---|---|
| bool | value | The new property value being set. |
Remarks
This method is invoked right before the value of HasEmg is changed.
OnHasEmgChanging(bool, bool)
Executes the logic for when HasEmg is changing.
Declaration
private void OnHasEmgChanging(bool oldValue, bool newValue)
Parameters
| Type | Name | Description |
|---|---|---|
| bool | oldValue | The previous property value that is being replaced. |
| bool | newValue | The new property value being set. |
Remarks
This method is invoked right before the value of HasEmg is changed.
OnHasImuChanged(bool)
Executes the logic for when HasImu just changed.
Declaration
private void OnHasImuChanged(bool value)
Parameters
| Type | Name | Description |
|---|---|---|
| bool | value | The new property value that was set. |
Remarks
This method is invoked right after the value of HasImu is changed.
OnHasImuChanged(bool, bool)
Executes the logic for when HasImu just changed.
Declaration
private void OnHasImuChanged(bool oldValue, bool newValue)
Parameters
| Type | Name | Description |
|---|---|---|
| bool | oldValue | The previous property value that was replaced. |
| bool | newValue | The new property value that was set. |
Remarks
This method is invoked right after the value of HasImu is changed.
OnHasImuChanging(bool)
Executes the logic for when HasImu is changing.
Declaration
private void OnHasImuChanging(bool value)
Parameters
| Type | Name | Description |
|---|---|---|
| bool | value | The new property value being set. |
Remarks
This method is invoked right before the value of HasImu is changed.
OnHasImuChanging(bool, bool)
Executes the logic for when HasImu is changing.
Declaration
private void OnHasImuChanging(bool oldValue, bool newValue)
Parameters
| Type | Name | Description |
|---|---|---|
| bool | oldValue | The previous property value that is being replaced. |
| bool | newValue | The new property value being set. |
Remarks
This method is invoked right before the value of HasImu is changed.
OnReceive(object, object)
Called by the upstream timer block. In vector mode, publishes the most recently decoded sample to downstream blocks and scope monitors.
Declaration
protected override void OnReceive(object sender, object tNow)
Parameters
| Type | Name | Description |
|---|---|---|
| object | sender | The timer block that triggered this callback. |
| object | tNow | The current pipeline timestamp. |
Overrides
Remarks
This method is a no-op in matrix mode, where data is published directly from PublishMatrixBatch(byte[], int).
PublishMatrixBatch(byte[], int)
Decodes a batch of consecutive samples into a MathNet.Numerics.LinearAlgebra.Matrix<T> and publishes it downstream (matrix / push mode).
Declaration
private void PublishMatrixBatch(byte[] buf, int batchOffset)
Parameters
| Type | Name | Description |
|---|---|---|
| byte[] | buf | The accumulation buffer containing raw bytes. |
| int | batchOffset | Byte offset of the first sample in the batch. |
ReaderLoop()
Background thread loop that continuously reads from the TCP stream, accumulates bytes into complete frames, and dispatches decoded data to the pipeline.
Declaration
private void ReaderLoop()
Remarks
The SyncStation sends data in bursts of SubSamplingRate (18) samples. TCP may fragment or coalesce these bursts, so the reader uses an accumulation buffer and processes as many complete frames as are available after each read.
SendConfiguration()
Builds and sends the SyncStation configuration packet over the open TCP stream.
Declaration
private void SendConfiguration()
Remarks
Packet format per OT Bioelettronica SyncStation protocol v2.8:
-
START byte A:
[0][REC_ON=0][SIZE4..SIZE0][GO=1]where SIZE is the count of CONTROL bytes that follow. -
CONTROL bytes (one per enabled probe):